Accuracy of QT measurement in the EASI-derived 12-lead ECG


Abstract:

The purpose of this study is to assess QT interval measurements from the EASI 12-lead EGG as compared to the standard 12-lead ECG. The QT interval was automatically measured in simultaneously recorded standard and EASI 12-lead ECGs, using a validated wavelet-based delineator. The agreement between the two sets of measurements was quantified both on a lead-by-lead basis and a multilead basis. The results show an acceptable agreement between QT measurements in the two lead systems, with correlation coefficients (CC) 0.91-0.98 depending on the lead. Standard deviations range from 9.2 ms to 26.4 ms depending on the selected lead. In a few patients large inter-system differences were found, mainly due to different T wave morphologies. Using a multilead delineation, QT intervals were considerably more stable than single-lead measurements and resulted in a much better agreement between the two lead systems (CC: 0.98, QT difference: 1.1 ms ± 9.8 ms). Thus, EASI-derived 12 lead ECG may be used for reliable QT monitoring when the multilead delineation approach is adopted. © 2006 IEEE.

Año de publicación:

2006

Keywords:

    Fuente:

    scopusscopus

    Tipo de documento:

    Conference Object

    Estado:

    Acceso restringido

    Áreas de conocimiento:

    • Enfermedad cardiovascular

    Áreas temáticas:

    • Medicina y salud
    • Enfermedades