Doped SiO2 telecommunication fibre as a 1-D detector for radiation therapy dosimetry
Abstract:
Present studies concern Ge-doped SiO2 telecommunication fibre as a high spatial resolution 1-D thermoluminescence (TL) system for radiotherapeutic dosimetry. Using tube xray bremsstrahlung sources operating at kilovoltage energies, these fibres have been shown to offer linear response, from < 1Gy up to in excess of 30 Gy. Measurement of the photoelectron dose enhancement resulting from use of a moderately high atomic number medium (iodinated contrast media) demonstrates the fibres to have the local dose sensitivity required of interface dosimetry. In PMMA, the TL yield is ~60% greater in the presence of iodine than in its absence.