ECG-Derived Respiratory Rate in Atrial Fibrillation


Abstract:

Objective: The present study addresses the problem of estimating the respiratory rate from the morphological ECG variations in the presence of atrial fibrillatory waves (f-waves). The significance of performing f-wave suppression before respiratory rate estimation is investigated. Methods: The performance of a novel approach to ECG-derived respiration, named 'slope range' (SR) and designed particularly for operation in atrial fibrillation (AF), is compared to that of two well-known methods based on either R-wave angle (RA) or QRS loop rotation angle (LA). A novel rule is proposed for spectral peak selection in respiratory rate estimation. The suppression of f-waves is accomplished using signal-and noise-dependent QRS weighted averaging. The performance evaluation embraces real as well as simulated ECG signals acquired from patients with persistent AF; the estimation error of the respiratory rate is determined for both types of signals. Results: Using real ECG signals and reference respiratory signals, rate estimation without f-wave suppression resulted in a median error of 0.015 ± 0.021 Hz and 0.019 ± 0.025 Hz for SR and RA, respectively, whereas LA with f-wave suppression resulted in 0.034 ± 0.039 Hz. Using simulated signals, the results also demonstrate that f-wave suppression is superfluous for SR and RA, whereas it is essential for LA. Conclusion: The results show that SR offers the best performance as well as computational simplicity since f-wave suppression is not needed. Significance: The respiratory rate can be robustly estimated from the ECG in the presence of AF.

Año de publicación:

2020

Keywords:

  • f-wave suppression
  • Respiratory rate
  • ECG-derived respiration
  • Atrial Fibrillation

Fuente:

scopusscopus

Tipo de documento:

Article

Estado:

Acceso restringido

Áreas de conocimiento:

  • Enfermedad cardiovascular

Áreas temáticas:

  • Medicina y salud