Analytical and clinical comparison of Viasure (CerTest Biotec) and 2019-nCoV CDC (IDT) RT-qPCR kits for SARS-CoV2 diagnosis.
Abstract:
Background: Several RT-qPCR kits are available for SARS-CoV-2 diagnosis, some of them with Emergency Use Authorization (EUA) by FDA, but most of them lacking of proper evaluation studies due to covid19 emergency. Objective: We evaluated Viasure RT-qPCR kit (CerTest Biotec, Spain) for SARS-CoV-2 diagnosis using FDA EUA 2019-nCoV CDC kit (IDT, USA) as a gold standard. Results: Although we found the lack of RNA quality control probe as the main limitation for the Viasure kit, the sensitivity was 91.9% and the specificity was 100%. The limit of detection (LOD) was 2000 copies/mL and 1000 copies/mL for Viasure and IDT kits, respectively. Conclusions: Viasure RT-qPCR kit is a reliable tool for SARS-CoV-2 diagnosis but improvement of an alternative RT-qPCR reaction for RNA extraction quality control as RNaseP is recommended.
Año de publicación:
2021
Keywords:
Fuente:


Tipo de documento:
Article
Estado:
Acceso abierto
Áreas de conocimiento:
- Microbiología
Áreas temáticas:
- Farmacología y terapéutica
- Enfermedades
- Medicina y salud