Influence of Photoplethysmogram Signal Quality on Pulse Arrival Time during Polysomnography


Abstract:

Intervals of low-quality photoplethysmogram (PPG) signals might lead to significant inaccuracies in estimation of pulse arrival time (PAT) during polysomnography (PSG) studies. While PSG is considered to be a “gold standard” test for diagnosing obstructive sleep apnea (OSA), it also enables tracking apnea-related nocturnal blood pressure fluctuations correlated with PAT. Since the electrocardiogram (ECG) is recorded synchronously with the PPG during PSG, it makes sense to use the ECG signal for PPG signal-quality assessment. (1) Objective: to develop a PPG signal-quality assessment algorithm for robust PAT estimation, and investigate the influence of signal quality on PAT during various sleep stages and events such as OSA. (2) Approach: the proposed algorithm uses R and T waves from the ECG to determine approximate locations of PPG pulse onsets. The MESA database of 2055 PSG recordings was used for this study. (3) Results: the proportions of high-quality PPG were significantly lower in apnea-related oxygen desaturation (matched-pairs (Formula presented.) = 0.88 and (Formula presented.) = 0.97, compared to OSA and hypopnea, respectively, when p < 0.001) and arousal ((Formula presented.) = 0.93 and (Formula presented.) = 0.98, when p < 0.001) than in apnea events. The significantly large effect size of interquartile ranges of PAT distributions was between low- and high-quality PPG (p < 0.001, (Formula presented.) = 0.98), and regular and irregular pulse waves (p < 0.001, (Formula presented.) = 0.74), whereas a lower quality of the PPG signal was found to be associated with a higher interquartile range of PAT across all subjects. Suggested PPG signal quality-based PAT evaluation reduced deviations (e.g., (Formula presented.) = 0.97, (Formula presented.) = 0.97, (Formula presented.) = 0.99 in hypopnea, oxygen desaturation, and arousal stages, respectively, when p < 0.001) and allowed obtaining statistically larger differences between different sleep stages and events. (4) Significance: the implemented algorithm has the potential to increase the robustness of PAT estimation in PSG studies related to nocturnal blood pressure monitoring.

Año de publicación:

2023

Keywords:

  • PAT
  • hypopnea
  • irregular heart rhythm
  • electrocardiogram
  • pulse onset
  • R wave
  • SpO2
  • photoplethysmogram
  • obstructive sleep apnea
  • T wave
  • polysomnography

Fuente:

scopusscopus

Tipo de documento:

Article

Estado:

Acceso abierto

Áreas de conocimiento:

  • Fisiología

Áreas temáticas:

  • Fisiología humana
  • Enfermedades
  • Medicina y salud