Mathematical model of premature baby thermoregulation and infant incubator dynamics


Abstract:

A thermodynamic mathematical model of premature infants placed in an infant incubator was developed to improve the control algorithm of the incubator. The validity of the baby model was tested in 5 healthy premature newborns. The difference between experimental data from direct and indirect calorimetry and the model results averaged 0.1±0.1, 0.2±0.1 and 0.2±0.2 (mean = SE) °C for the rectum, trunk and head temperatures, respectively. The incubator model showed steady state error less than ±0.1 C for air and internal wall temperatures in the air servocontrol mode. Results indicated that the model is a good tool for designing improved control algorithms of incubators.

Año de publicación:

1996

Keywords:

    Fuente:

    scopusscopus

    Tipo de documento:

    Conference Object

    Estado:

    Acceso restringido

    Áreas de conocimiento:

    • Modelo matemático

    Áreas temáticas:

    • Ciencias de la computación
    • Ginecología, obstetricia, pediatría, geriatría
    • Fisiología humana