QCM modified with FAU zeolite nanostructures for analysis of temperature induced adsorbed mass changes


Abstract:

In this work, the integration of various elements was realized. This allowed the development of a measurement system for quantitative determination of the electrical response of the faujasite-modified QCM, which was caused by a change in mass during the adsorption and desorption processes. The nanostructured faujasite film was grown on the QCM surface, and water was used as a ”model molecule” to test the ability of a measurement system to determine the change in mass associated with adsorption and desorption. It was shown that the sensitivity of the system is equal to 17 ng/Hz. The obtained sensitivity is sufficient to observe not only the adsorption and desorption process but even climate changes in the surrounding atmosphere.

Año de publicación:

2021

Keywords:

  • Zeolite
  • Water adsorption
  • QCM
  • Gas sensing

Fuente:

scopusscopus

Tipo de documento:

Article

Estado:

Acceso restringido

Áreas de conocimiento:

  • Nanostructura
  • Ciencia de materiales
  • Ciencia de materiales

Áreas temáticas:

  • Química analítica
  • Ingeniería química