Automatic Calibration for Residential Water Meters by Using Artificial Vision


Abstract:

The present work addresses the problem of automated calibration system for residential water meters, by using artificial vision. The project consists of a closed water flow circuit powered by a low-power pump; the data from water meter is taken by the computer using a USB camera; a calculation is made based on the time it takes to fully rotate the smaller-scale needle of the meter to determine the water flow in real time. At the same time, the actual flow data of the reference standard element, which is a rotameter, is obtained by a second camera. These values are compared to calculate an error that determines the adjustment action on the water meter.

Año de publicación:

2020

Keywords:

  • artificial vision
  • Automatic calibration
  • LabVIEW
  • Water meter

Fuente:

googlegoogle
scopusscopus

Tipo de documento:

Conference Object

Estado:

Acceso restringido

Áreas de conocimiento:

  • Automatización
  • Visión por computadora
  • Recursos hídricos

Áreas temáticas:

  • Física aplicada
  • Instrumentos de precisión y otros dispositivos
  • Métodos informáticos especiales