Energy resolution enhancement of mercuric iodide detectors
Abstract:
A pulse processing technique has been developed which improves the gamma-ray energy resolution of mercuric iodide detectors. The technique employs a fast (100 ns) and a slow (6.4 μs) pulse height analysis to correct for signal variations due to variations in charge trapping. The capabilities of the technique for energy resolution enhancement are discussed as well as the utility of the technique for examining the trapping characteristics of individual detectors. We have achieved an energy resolution of 2.6% FWHM at 662 keV with an acceptance efficiency of 100% from a mercuric iodide detector which gives 8.3% FWHM using standard techniques.
Año de publicación:
1984
Keywords:
Fuente:

Tipo de documento:
Other
Estado:
Acceso abierto
Áreas de conocimiento:
- Ciencia de materiales
Áreas temáticas:
- Física aplicada
- Electricidad y electrónica